How to Measure Thermocouples
Temperature and Thermocouple Overview
Temperature is the average kinetic energy of the particles in the object sample measurement, the
standard unit is "degree." Temperatures can be measured in different ways, and the cost and
accuracy of the measurements can vary. Thermocouples are one of the most common temperature sensors
, because thermocouples are relatively inexpensive and accurate, and have a relatively wide
measurement range.
Whenever two different metal contacts, the contact point assembly produces a lower no-load voltage
as a function of temperature, which is the thermoelectric effect. The temperature difference is
the Seebeck voltage, to 1821 found the phenomenon of physicist Thomas Seebeck named. This voltage
is non-linear with respect to temperature, but can be approximated as being linear for small
changes in temperature, or:

Where ΔV is the voltage change, S is the Seebeck coefficient, and ΔT is the temperature change.
Thermocouples are available in a variety of types and are specified in larger letters according to
the American National Standards Institute (ANSI) convention. For example, a J-type thermocouple
consists of an iron conductor and a copper-nickel alloy conductor. Other types of thermocouples
include B, E, K, N, R, S, and T.
How to Measure Thermocouples
background knowledge
For a better understanding of how thermocouple measurements are made, it is important to understand
how thermocouples work. The first part of this document will explain the basic principles of
thermocouples, the later part of the thermocouple will explain how to achieve the connection
between the instrument and how the temperature measurement.
The thermocouple Seebeck voltage, if connected directly to the measuring system, is connected to
the measuring system and generates an additional thermocouple. This can not be measured simply by
connecting it to a voltmeter or other measuring system.

Figure 1. J-type thermocouple
As shown in Figure 1, the circuit using the J-type thermocouple on the candle temperature
measurement. Two thermocouple lines are connected to the copper terminals of the data acquisition
device. Note that there are three metal connections in this circuit - J1, J2 and J3. J1 is the
thermocouple measurement point, producing a Seebeck voltage proportional to the temperature of the
candle. In addition, J2 and J3 each have their own Seebeck coefficient, and in the data
acquisition terminal will produce a temperature proportional to the temperature difference voltage
, known as the cold-side voltage. In order to determine the voltage component of J1, it is
necessary to know the temperature of J2 and J3 contacts and know the relationship between the
contact voltage and the temperature. Thus, the voltage of the J1 junction can be obtained by
subtracting the J2 and J3 parasitic junction voltage components from the measured voltage.
Thermocouples require a specific temperature reference to compensate for the cold side of the
error generated. The most commonly prescribed method is to use a temperature sensor that can be
directly read to obtain the reference junction temperature, minus the parasitic terminal voltage
component. This process, known as cold junction compensation, can simplify the calculation of cold
junction compensation using the characteristics of some thermocouples.
By using the thermocouple law of the metal transition layer and other assumptions, we can see that
the measurement of the voltage data acquisition system depends only on the type of thermocouple,
the voltage at the measurement terminal, and the cold junction temperature. The measured voltage
is independent of the voltage components of the measuring leads and the cold terminals J2, J3.

Figure 2. Metallic transition layer thermocouple law
Consider the circuit in Figure 3. This circuit is similar to the circuit described in Figure 1
above, except that a short piece of copper-nickel alloy wire is inserted before the J3 junction.
All contacts are at the same temperature. Assuming that the J3 and J4 junction temperatures are
the same, the metal transition layer thermocouple law states that the circuit in Figure 3 is
electrically the same as the circuit in Figure 1. Therefore, any of the results of the circuit of
Fig. 3 can be applied to the circuit shown in Fig.

Figure 3. Inserting an additional conductor in an isothermal environment
Figure 3, J2 and J4 contacts belong to the same type (copper-nickel alloy); because the two are in
an isothermal environment, J2 and J4 is the same temperature. Because of the direction of the
current in the circuit, J4 produces a Seebeck positive voltage, J2 produces a Seebeck negative
voltage. Therefore, the contact offset the influence of each other, measuring the total voltage is
zero. J1 and J3 contacts are iron-copper-nickel alloy contacts. But their temperatures may be
different, as they may not be in an isothermal environment. Because they are in different
temperature environments, J1 and J3 contacts can produce Seebeck voltage, but the size is different
. In order to compensate for the cold junction J3, the temperature is measured and the applied
voltage is subtracted from the thermocouple measurement.
Using the VJx (Ty) symbol to represent the voltage generated by the Jx junction at Ty temperature,
the general thermocouple problem is simplified as follows:
VMEAS = VJ1(TTC ) + VJ3(Tref ) (2)
Where, VMEAS data acquisition system that the measured voltage value, TTC J1 junction thermocouple
temperature, Tref said reference side of the temperature.
Note that in the formula (2), VJx (Ty) represents the voltage generated in a Ty temperature
environment with respect to a certain reference temperature. Equation 2 holds as long as VJ1 and
VJ3 are temperature functions related to the same reference temperature. For example, a NIST
thermocouple reference table as described above is generated with the reference end held at 0
degrees Celsius.
Because J3 and J1 are of the same type, but produce relative voltages, VJ3 (Tref) = -VJ1 (Tref).
Also, because VJ1 is the voltage generated by the thermocouple type test, the voltage can be
renamed to VTC. Therefore, 2-type can be rewritten as follows:
VMEAS = VTC (TTC ) - VTC (Tref ) (3)
Therefore, by measuring VMEAS and Tref know the relationship between the thermocouple voltage and
the temperature, you can determine the temperature of the thermocouple measurement side.
There are two technologies to achieve cold junction compensation - hardware compensation and
software compensation. Both techniques require the use of a direct-read sensor to obtain the
reference-side temperature. The direct-reading sensor has an input that is only determined by the
temperature of the measuring point. Semiconductor sensors, thermistors and RTDs are commonly used
to measure the temperature of the reference side of the instrument.
Using hardware compensation, a variable voltage source can be inserted into the circuit to remove
the parasitic thermoelectric voltage. The variable voltage source generates a compensation voltage
according to the ambient temperature, which is added to the correction voltage to cancel the
unwanted temperature difference signal. When these parasitic signals are removed, the only signal
measured by the data acquisition system is the voltage measured from the thermocouple measurement
. In the case of hardware compensation, the temperature of the data acquisition system terminal is
uncorrelated because the parasitic thermocouple voltage in the data acquisition system has been
canceled. The main disadvantage of hardware compensation is that each thermocouple must have a
separate compensation circuit can be added to correct the compensation voltage, which will greatly
increase the cost of the circuit. Under normal circumstances, the hardware compensation is not on
the accuracy and software compensation.
Or you can choose to use the software for cold junction compensation. After using the
direct-reading sensor to measure the reference-side temperature, the software can append a suitable
voltage value to the measured voltage to eliminate the influence of the cold-side voltage. Recall
(3) that the measured voltage VMEAS is equal to (thermocouple) measuring the voltage difference
between the termination point and cold junction.
The thermocouple output voltage is highly non-linear. The Seebeck coefficient varies due to three
or more factors in the operating temperature range of some thermocouples. Therefore, you must use
a polynomial to model the voltage versus temperature curve in a thermocouple or use the look-up
table method.
Connect the thermocouple to the instrument
This section uses the NI cDAQ-9172 backplane and the NI 9211 C-series thermocouple module as an
example. A similar procedure is applicable for connecting thermocouples to different instruments
(see Figure 4).
Required equipment:
DAQ-9172 8-slot high-speed USB backplane for NI CompactDAQ
NI 9211 four-channel, 14Sa / s, 24-bit, ∓ 80 millivolt thermocouple input module
J type thermocouple

Figure 4. NI CompactDAQ system
The NI 9211 features a 10-pole, detachable screw terminal connector that provides connections for
up to four thermocouple input channels. Each channel has a connection point for the positive
thermocouple, TC +, and a connection to the negative terminal, TC-. The NI 9211 also has a common
wiring point, COM. Normally this port is internally connected to the reference ground of the module
. Figure 5 shows the wiring assignments for each channel, and Figure 6 shows the wiring diagram.

Figure 5. Terminal assignment

Figure 6. Connection diagram
View your measurements: NI LabVIEW
Now that the thermocouple is connected to the test equipment, you can use LabVIEW graphical
programming software to transfer the data to the computer for visualization and analysis.
Figure 7 shows an example of displaying the measured temperature data in the LabVIEW programming
environment.

Add: No. 16, Road 3, Sanjiang Industry Zone, Shengzhou, Zhejiang, China
Contact:Eric Xie
Email: sale@benoelectric.com
Mob: +861598895936
Skype: ericxie123@outlook.com
Wechat & Whatsapp: +8615988295936
Website:www.benoindustry.com
